Maintenance
65
Reassemble in reverse order
NOTICE
It is absolutely important to hold the brush holder (8) and not the bar (5), when unscrewing the
heat shield (9), otherwise the rods (6) will bend.
6.8 Dust trap cleaning
CAUTION
C23.0076
Scalding/burns
Hot furnace / combustion tube / analyzer parts
Parts of the analyzer can be very hot.
Use heat protecting gloves.
WARNUNG
W13.0023
High temperature / Burning of the skin
Hot combustion tube
When the analyzer (HTF, resistance furnace) is switched on again after
switching off, the adjusted set point is driven immediately. Please
check this!
6.8.1 Dust trap cleaning-Induction furnace
The dust from the furnace is trapped in the metal dust filter (2). The dust filter needs to be
cleaned after about 100 analyses.
Make sure that the filter is absolutely dry after ultrasonic cleaning. The porosity of the filter is 10
microns only and it is impossible to recognize whether the inside of the filter is dry, by simply
looking into it. Therefore it is advisable to have a second filter, in order to use it while drying the
other one. The filter can be dried with hot air. The replacement of the dust filter (2) takes only
about 5 seconds.

Table of Contents

Other manuals for ELTRA CS-2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ELTRA CS-2000 and is the answer not in the manual?

ELTRA CS-2000 Specifications

General IconGeneral
BrandELTRA
ModelCS-2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals