Maintenance
47
Fig.
23
: Replacement of moisture trap
1. Turn both screws (1) counter clockwise, until the ring (3) touches the part (2).
2. Lift the glass tube (5) upwards, tilt it to the side and pull downwards.
3. Replace the magnesium perchlorate (Anhydrone). See chapter reagent tube replacing
4. Install the glass tube (5) in reverse order after greasing the lower O-ring.
5. Turn both screws (1) clockwise until the O-ring (4) is properly pressed on the
inner surface of the glass tube (5).
Remark
The O-ring imprint on the whole circle length of the glass tube should be of about 2 mm width.
6.3 Reagent tubes filling
6.3.1 Chemicals
CAUTION
C13.0076
Scalding/burns
Hot furnace / combustion tube / analyzer parts
Parts of the analyzer can be very hot.
Use heat protecting gloves.
WARNING
W9.0017
Danger of toxication and personal injuries
Some chemicals may cause a fatal toxication or dangerous skin corrosion.
Refer to the material safety data sheet of the used substances.
Never eat or drink close to the chemical substances.

Table of Contents

Other manuals for ELTRA CS-2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ELTRA CS-2000 and is the answer not in the manual?

ELTRA CS-2000 Specifications

General IconGeneral
BrandELTRA
ModelCS-2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals