Maintenance
51
The furnace of the analyzer is fitted with an automatic furnace cleaning system with thermal
shield .
Proceed as follows to replace the furnace cleaning brush with thermal shield:
1. Open the furnace.
As a result, the springs at the top of the furnace are released and the knurled nuts can be
easily unscrewed.
2. Turn the main switch on the analyzer to Position 0 and pull out the power plug.
3. Wait until the furnace of the analyzer has cooled down sufficiently.
The following diagram shows the mounting bolts for the top furnace lining at the top and at the
front of the machine:
Fig. 14: Removing the top furnace lining
4. Unscrew the Allen screws (1), (3) and (4) at the top and front of the furnace lining.
5. Remove the furnace lining (2).
1
3
2
4