3410 Series Gas Ultrasonic Meters Operations Manual Section 6: Archive logs
3-9000-777 Rev E January 2018
System log 113
6.2.1 System log
The Daniel 3410 Series Gas Ultrasonic Flow Meter meter logs all system messages in the system
log. The meter can store up to 3000 system records. The user can select whether old, unread
records can be overwritten by new records when the log becomes full via the data point
DoOverwriteUnreadSystemLog. This point can be modified using the Daniel MeterLink
Tools|Edit/Compare Configuration screen. The default is to overwrite old, unread records. Refer
to “Options for reading audit, alarm, and/or system log records” for information on reading
records and marking records as read. The data point
IsSystemLogFull indicates whether or not
the system log is full and cannot overwrite old, unread records.
Repetitive system messages are prevented from filling up the system log. When a particular
system message occurs 3 times within 60 seconds, that system message is suppressed until that
message’s frequency drops below the 3 times-per-60 sec rate.
The system log records indicate
when a system message suppression is started and ended.
6.2.2 Reading log records
Daniel 3410 Series Ultrasonic Flow Meter log records are read using the Daniel MeterLink
Logs/Reports|Meter Archive Logs screen. The are three log groups:
daily
hourly
event (audit, alarm, and system logs)
Select the desired log group(s) via the
Collect daily log/Collect hourly log/Collect event log check
boxes. If the event group is selected, the audit, alarm, and system logs are individually
selectable. Whenever any log is collected, the current meter configuration is also collected.
6.2.3 Options for reading daily and/or hourly log records
The options for reading daily and hourly log records are the same. The Daniel MeterLink
Logs/Reports|Meter Archive Logs screen indicates the number of daily records that are available
to read. Select the log type(s) to be collected via the Collect daily log and/or Collect hourly log
check boxes. Select whether to collect all log records or the last specified number of daily
records. Also, select whether to collect all log data or just the data points required by the API
Chapter 21 standard. Tab le 6-1 lists the daily log data points and Ta bl e 6 -2 lists the hourly log
data points. For both tables, data points required by the API Chapter 21 standard are marked
with an asterisk (*).
6.2.4 Options for reading audit, alarm, and/or system log records
The options for reading audit, alarm, and system log records are the same. The Daniel MeterLink
Logs/Reports|Meter Archive Logs screen indicates the number of records available for each log
type. Select whether to collect either all the records or just the last specified number of daily
records for the selected log type.

Table of Contents

Other manuals for Emerson Daniel 3410 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Daniel 3410 Series and is the answer not in the manual?

Emerson Daniel 3410 Series Specifications

General IconGeneral
BrandEmerson
ModelDaniel 3410 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals