3410 Series Gas Ultrasonic Meters Operations Manual Section 6: Archive logs
3-9000-777 Rev E January 2018
Collecting Logs 115
6.2.6 Collecting Logs
To collect Maintenance and Trend Logs for meter diagnostics,
1. Establish communication with the meter via Daniel MeterLink.
2. From Daniel MeterLink main window, select
Logs/Reports|Maintenance Logs and Reports.
The Maintenance Logs and Reports dialog
displays.
Figure 6-2 Maintenance Logs
3. Set the log collection Duration (mins) (2 minutes is the default value which is normally
adequate to capture a good snapshot of the meter's current diagnostic parameters). To
get a good representation of the meter performance, it is recommended that there be
at least 30 records (lines of data). The actual number of records collected are a function
of the type of communication (Serial or Ethernet), Stack Size (if stacking is turned on),
and the Collection speed selected. The duration value can be changed by clicking the
number in the display/edit box and entering a new value.