3410 Series Gas Ultrasonic Meters Operations Manual Section 5: Calibrate and configure the meter
3-9000-777 Rev E January 2018
Dry calibration 41
The meter provides separate dry calibration coefficients for each flow direction as listed in the
table below.
Table 5-1 Data Points for dry calibration
Daniel MeterLink
Display Name
Data Points, Options and Guidelines
Forward A0 Data points affected:
• FwdA0
Options:
• Enter a value (m/s or ft/s) within the range [-1,1 m/s].
Guidelines:
• This value should only be modified under direction of Daniel personnel.
Forward A1 Data points affected:
• FwdA1
Options:
• Enter a value (dimensionless) within the range [0.95,1.05].
Guidelines:
• This value should only be modified under direction of Daniel personnel.
Forward A2 Data points affected:
• FwdA2
Options:
• Enter a value (s/m or s/ft) within the range [-0.1,0.1 s/m].
Guidelines:
• This value should only be modified under direction of Daniel personnel.
Forward A3 Data points affected:
• FwdA3
Options:
• Enter a value (s
2
/m
2
or s
2
/ft
2
) within the range [-0.1,0.1 s
2
/m
2
].
Guidelines:
• This value should only be modified under direction of Daniel personnel.
Reverse A0 Data points affected:
• RevA0
Options:
• Enter a value (m/s or ft/s) within the range [-1,1 m/s].
Guidelines:
• This value should only be modified under direction of Daniel personnel.
Reverse A1 Data points affected:
• RevA1
Options:
• Enter a value (dimensionless) within the range [0.95,1.05].
Guidelines:
• This value should only be modified under direction of Daniel personnel.