146 COMMUNICATIONS TROUBLESHOOTING
Troubleshooting comm., mech., and elec. Daniel 3410 Series Gas Ultrasonic Meter Operations Manual
January 2018 3-9000-777 Rev E
If you get the following:
Ethernet adapter Local Area Connection 1:
IP Address: 0.0.0.0
then the PC has not yet received an IP address from the DHCP server and you should wait (up to 30 seconds)
to receive an IP address before attempting to connect to the meter. If after 30 seconds the PC has not
received an IP address from the meter DHCP server or the IP address shown above (from ipconfig) is different
from the range of 192.168.135.35 through 192.168.135.44, verify that the PC is configured to receive its IP
address automatically (via DHCP).
To ensure connection to the meter from the PC, at the DOS prompt type:
ping 192.168.135.100 <enter>
If the meter is reachable, then you will see a message like:
Pinging 192.168.135.100 with 32 bytes of data:
Reply from 192.168.135.100: bytes=32 time < 10ms TTL=64
etc.
If the meter is not reachable, then you will see something like:
Pinging 192.168.135.100 with 32 bytes of data:
Request Timed Out
etc.
Q3. How do I connect to multiple meters via Ethernet when they are on the same LAN?
A3. Before connecting multiple meters via Ethernet on a LAN, each meter must be configured with a unique user-
specified IP address (following the initial communication quick start instructions in Section 2.1). Contact
your IT department for valid IP addresses for your LAN and Gateway address if required. Once a meter's IP
address is configured, the meter may be connected to the intranet LAN and accessed using that IP address.
3410 Series meters connected to an intranet LAN should not have their DHCP servers enabled.
Q4. How do I connect to multiple meters via Ethernet when they are on the same hub but not connected to an
intranet LAN?
A4. The PC may receive its IP address from an external DHCP server; in this case, one and only one meter must
have its DHCP server enabled. This DHCP server will serve up to 10 IP addresses to PCs attempting to talk to all
meters on the hub.
Before connecting multiple meters via Ethernet on a hub, each meter must be configured with a unique user-
specified IP address (following the initial communication quick start instructions in Section 2.1). Assign each
meter on the hub a unique IP address within the range 192.168.135.150 through 192.168.135.254. The
Gateway address for each meter may be left unconfigured as 0.0.0.0. Once a meter's IP address is configured,
the meter may be connected to the hub and accessed using that IP address.

Table of Contents

Other manuals for Emerson Daniel 3410 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Daniel 3410 Series and is the answer not in the manual?

Emerson Daniel 3410 Series Specifications

General IconGeneral
BrandEmerson
ModelDaniel 3410 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals