Rosemount Model 1195/ProPlate/Mass ProPlate
7-6
XMTR INFO
A non-volatile memory fault has been detected in the flowmeter
memory by the memory verification routine. The memory fault is in a
location containing flowmeter information. To correct this problem, use
a Model 275 HART Communicator to interrogate and reconfigure the
appropriate portion of the flowmeter memory. This warning does not
effect the flowmeter operation. Contact Dieterich Standard Inc. at
303-530-9600 if you need assistance.
Operation
Normal operation messages appear on the LCD meter to confirm
actions or inform you of flowmeter status. Operation messages are
displayed with other flowmeter information, and warrant no action to
correct or alter the flowmeter settings.
ZERO PASS
The zero value, set with the local zero adjustment button, has been
accepted by the flowmeter, and the output should change to 4 mA.
ZERO FAIL
The zero value, set with the local zero adjustment button, exceeds the
maximum rangedown allowed for a particular range, or the pressure
sensed by the flowmeter exceeds the sensor limits.
SPAN PASS
The span value, set with the local span adjustment button, has been
accepted by the flowmeter, and the output should change to 20 mA.
SPAN FAIL
The span value, set with the local span adjustment button, exceeds the
maximum rangedown allowed for a particular range, or the pressure
sensed by the flowmeter exceeds the sensor limits.
LOCAL DSBLD
This message appears during reranging with the integral zero and span
buttons and indicates that the flowmeter local zero and span
adjustments have been disabled. The adjustments may have been
disabled by the flowmeter security jumper on the flowmeter circuit
board or through software commands from the Model 275. Refer to
Flowmeter Security on page 5-3 for information on the position of the
security jumper.
WRITE PROTCT
This message appears if you attempt to change the flowmeter
configuration data while the security jumper is in the ON position.
See Flowmeter Security on page 5-3 for more information about the
security jumper.