EasyManua.ls Logo

Emerson Rosemount 1195 - Valves and Fittings; Impulse Piping

Emerson Rosemount 1195
229 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rosemount Model 1195/ProPlate/Mass ProPlate
2-20
Perform the following procedure to install adapters to the 1195 Integral
Orifice process connections:
1. Place non-threaded side facing 1195 process connections. This will
leave thread side facing transmitter.
2. Install and tighten bolts and tighten bolts and nuts. Refer to
Mounting Bolts on page 2-26 for torque specifications.
When compressed, Teflon
®
(PTFE) O-rings tend to cold flow, which aids
in their sealing capabilities. Whenever you remove flanges or adapters,
visually inspect the Teflon O-rings. Replace them if there are any signs
of damage, such as nicks or cuts. If they are undamaged, you may reuse
them. If the you replace the O-rings, re-torque the flange bolts after
installation to compensate for cold flow. Refer to the process sensor body
reassembly procedure in Section 9: Troubleshooting.
Valves and Fittings Throughout the remote mounting process:
Use only valves and fittings rated for the service pipeline design
pressure and temperature as specified in Section 10:
Specifications and Reference Data.
Use a pipe thread sealant compound that is rated for use at the
service temperature and pressure for all valves and fittings.
Verify that all connections are tight and that all instrument
valves are fully closed.
Verify that the meter assembly is properly oriented for the
intended type of service: liquid, gas or steam (see Figures 2-23,
2-24, and 2-25).
Impulse Piping A remote mounted electronics is connected to the sensor by means of
impulse piping. Temperatures in excess of 450°F at the electronics
will damage electronics components; impulse piping allows service
flow temperatures to decrease to a point where the electronics is no
longer vulnerable.
Failure to install proper flange adapter O-rings can cause process leaks, which can
result in death or serious injury.
The flange adapters require a unique O-ring, as shown below.
Unique O-ring Grooves
Flange Adapter
O-ring

Table of Contents

Other manuals for Emerson Rosemount 1195

Related product manuals