Troubleshooting
March 2016
Troubleshooting
Detaching the meter adapter
The above instructions will provide access to the meter body. Use the following steps if it is
necessary to remove the coaxial cable:
1. Loosen and remove the two screws that hold the union onto the meter adapter and
pull the union away from the adapter.
2. Loosen and remove the sensor cable nut from the other end of the union.
3. Loosen and disconnect the conduit adapter or cable gland from the meter adapter.
Attaching the meter adapter
1. If you are using a conduit adapter or cable gland, slide it over the plain end of the
coaxial cable (the end without a ground wire).
2. Slide the meter adapter over the coaxial cable end.
3. Use a
5
/16-in. (8 mm) open end wrench to securely tighten the sensor cable nut onto
one end of the union.
4. Place the union onto the two screws extending out of the meter adapter and
tighten the two screws.
Connecting the coaxial cable at the meter body
1. Pull the sensor cable out of the support tube slightly and securely tighten the
sensor cable nut onto the union.
Note
Do not stretch the sensor cable over 1.5 inches (40 mm) beyond the top of the support
tube. Damage to the sensor may occur if the sensor cable is stressed.
2. Place the meter adapter into the top of the support tube and line up the screw
holes.
3. Use a hex wrench to turn the three adapter screws counterclockwise (outward) to
engage the support tube.
4. Tighten the conduit adapter or cable gland into the meter adapter.