Electronics Verification
March 2016
Reference Manual
00809-0100-4772, Rev FA
159
Electronics Verification
C.3.2 SI units
Example 1
In this application, an input frequency of 80 Hz represents a flow rate of 1,686.8 lpm.
Given
Q = flow rate
F = frequency: 80 Hz
K = K-factor (compensated): 10.772 (via fieldbus configuration tool)
C
x
= C
8
C
8
= 4.40287E-03 (from Table C-1 on page 156)
Other conditions
Fluid = water
Line size = 80 mm
Line pressure = 700 kPag
Operating temperature = 60 °C
Equation
Calculations
Example 2
In this application, an input frequency of 650 Hz represents a flow rate of 3,446.4 kg/hr.
Given
M = mass flow rate
F = frequency: 650.00 Hz
K = K-factor (compensated): 10.715 (via fieldbus configuration tool)
C
x
= C
16
C
16
= C
11
/ρ (from Table C-1 on page 156)
ρ = density: 4.169 lb/m
3
(operating)
Q
F
KC
x
×
-------------- -
=
Q
80
10.772 C
8
×
------------------------- -
=
Q
80
10.772 0.00440287
×
-------------------------------------------
=
Q 1,686.8 lpm=
Aotewell Ltd
www.aotewell.com
Industry Automation
HongKong|UK|China
sales@aotewell.com
+86-755-8660-6182

Table of Contents

Other manuals for Emerson Rosemount 8800D Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount 8800D Series and is the answer not in the manual?

Emerson Rosemount 8800D Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount 8800D Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals