2.7.10 FILTER_AUTO_ADJUST
Optimize Signal Processing based on the density of the process fluid. Select the incremental
value equal to or less than the process fluid density.
2.7.11 PRIMARY_VALUE_RANGE
Range of the flow measurement. The units match the XD_SCALE units from the AI Block
assigned to Flow
2.7.12 SENSOR_RANGE
Range the meter is capable of measuring. The maximum measurable flow rate the vortex
meter will measure
The units match the XD_SCALE units from the AI Block assigned to Flow
2.7.13 SECONDARY_VALUE
The shedding frequency measured by the vortex sensor in units of Hz
2.7.14 INSTALLATION_EFFECTS
Installation Effect correction factor. Use when the meter is installed in less than ideal piping
configuration. See Installation Effects document 00816-0100-3250 for details.
Valid range -1.5 to 1.5
2.7.15 PROCESS_DENSITY_RATIO
Density Ratio = Density at flowing P and T/Density at Standard P and T
The transmitter only uses density ratio when the flow output is in Standard or Normal units.
Examples of Standard unit: Standard Cubic Feet per Minute, SCFM
Example of Normal unit: Normal Cubic Meters per hour, NCMH
2.7.16 METER_DISPLAY
Data displayed locally on LCD display.
The following measurements can be displayed on the flowmeter LCD display:
Flow
Transducer Percent Range—Percent range of maximum accurate flow
Process Temperature—MTA electronics only
Process Density—MTA electronics only
Shedding Frequency
Electronics Temperature—MTA electronics only
Integrator Block Out—Totalized flow from Vortex meter Integrator Block