PlasmaQuant MS product family Specification
133
Ion Optics Ion optics system with ReflexION, a 3D focusing, 90 degree ion
mirror
Mass analyzer Low noise, double off-axis HD mass analyzer consisting of two
assemblies the fringe filter and the quadrupole filter (3 MHz). Self-
cleaning, maintenance free.
Detector AD Detector - All-digital detection system (10 MHz), with 11-orders
linear dynamic range and long life time
Signal Counts or counts/second
Mass range 3 to 260 amu
Resolution 0.5 to 1.2 amu, adjustable
(device with AMR option: for m/z > 230 amu resolution > 2 amu)
Supply voltage
Frequency
200-240 V AC ± 5 %
50/60 Hz
Line current 18 A max.
Typical power consumption 2700 VA
Max. power consumption 3700 VA
Overvoltage category II according to DIN EN 61010-1
Degree of contamination 2 according to DIN EN 61010-1
Protection class I
Protection type IP 20
Type of Gas Argon Helium Hydrogen Nitrogen* Oxygen*
Purity ≥ 4.6 (99.996 %)
5 ppm O
2
20 ppm N
2
4 ppm H
2
O
1 ppm H
2
≥ 4.6 (99.996 %) 4.6 (99.996 %) ≥ 4.6 (99.996 %) ≥ 4.6 (99.996 %)
Input Pressure 550-700 kPa
(80-100 psi)
regulated
150 kPa
(21.5 psi)
regulated
100 kPa
(14.5 psi)
regulated
125 kPa
(18.1 psi)
regulated
125 kPa
(18.1 psi)
regulated
Max. Input
Pressure
700 kPa (100 psi) 250 kPa
(36.3 psi)
170 kPa
(24.7 psi)
150 kPa
(21.5 psi)
150 kPa
(21.5 psi)
Flow Range 0-28 L/min 0-0.2 L/min 0-0.2 L/min 0-0.2 L/min 0-0.2 L/min
Typical flow
range
9-12.5 L/min
* for Nitrox option
Corrosion protection The device is corrosion-proof against the samples used in the
analysis
Operating
temperature
+15 to 25 °C, optimum +20 to 24 °C (as constant as possible during
measuring, maximum temperature drift ∆T
max
= 2 K/h, air condition
recommended)
Humidity 20 to 80 % at 20 °C (non-condensing)
Mass spectrometer
system
Power supply
Gas supply
Environmental conditions