Specification PlasmaQuant MS product family
134
Altitude 0 to 2000 m
Storage Temperature: +5 °C to +45 °C, use desiccant
For optimum analytical performance, it is recommended that positive air pressure be
maintained in the laboratory at all times.
Material Heat and corrosion resistant (recommended V2A steel)
Extraction output 3.0 m
3
/min (110 ft
3
/min) – 4.5 m
3
/min (160 ft
3
/min)
recommended value: 3.5 m
3
/min (125 ft
3
/min)
Adapter using
flexible aluminum or
plastic pipe
Internal pipe diameter ∅ 100 mm (4 in)
Length: < 5000 mm (15 ft)
Mass 186 kg
Dimensions
(W x D x H)
660 mm x 589 mm x 1131 mm
Leave a clearance of 30 cm on the right-side of the ICP-MS for opening the interface
door.
15.1.2 Control computer data
Processor 2.8 GHz, or better
Hard Disk Drive min. 80 GB hard disk drive storage capacity
RAM 4 GB of RAM
Graphic resolution
CD/DVD Drive 16x DVD RW+/- drive
USB Ports min. 2x USB 2.0 ports
Serial Ports min 1x
Multimedia Integrated video, audio, and network support
Operating System Windows 10 (32- or 64-bit), Windows 7 and 8.1 are supported
15.1.3 Water cooler data
Supply voltage
Frequency
200-240 V AC
50/60 Hz
Cooling Capacity Minimum of 1.8 kW
Water Temperature 20 °C ± 1 °C (68 °F ± 1.8 °F)
(temperature of the water supplied to the instrument)
Water Pressure 440 ± 40 kPa (64 ± 6 psi)
Pressure at the instrument water inlet