PlasmaQuant MS product family Installation and commissioning
37
The workbench surface must be wipeable, corrosion resistant and must not absorb
moisture.
It must be freely accessible from all sides. A distance of at least 200 mm to the next
wall must be kept. Alternatively, a heavy lift roll-top table can be used.
Due to the dissipated heat and possible noise it is recommended to place the water
cooler outside the laboratory. An extension of the cooling water tubing is permitted if
the minimum pressure and the flow volume are maintained. The surfaces of the
housing sides of the water cooler require a minimum distance of 15 cm from adjacent
objects, for unimpeded cooling air circulation.
The fore-line roughing pump is a floor-mounted pump, typically placed under or
behind the bench where the PlasmaQuant MS is located.
The pump can also be located remotely (up to 12 m away from the ICP-MS) for noise
reduction, and this is also recommended for Clean Room operation.
A bottle for receiving residue sample liquids and autosampler wash liquid is placed
underneath the bench.
Component Width
[mm]
Height
[mm]
Depth
[mm]
Mass
[kg]
On the workbench
PlasmaQuant MS 660 1131 589 186
Autosampler (ASPQ 3300) 285 510 490 15
Atosampler (Cetac ASX-560) 580 620 550 12
Outside the laboratory / or next to the workbench
Water-air cooler 460 703 735 92
Water-water cooler 360 590 470 33 (empty)
Fore-line roughing pump
Model SV40BI
Model XDS46i
330
340
380
400
500
480
43
48
Under the work table
Waste bottle 200 400