Installation and commissioning PlasmaQuant MS product family
42
4.3 Installing the PlasmaQuant MS
NOTICE
After transport, wait for 12 hours until the PlasmaQuant MS has reached room
temperature. Only start up the device after this period has elapsed. The sensitive
electronics may
otherwise be damaged by condensed water.
The PlasmaQuant MS is delivered directly to its final destination by a transportation
company. The delivery by this company requires the presence of a person responsible
for instrument installation.
It is imperative that all persons designated to operate the device are present during
the briefing by the service technician.
Check that all requirements are met at the installation location (see the section
"Installation conditions" p. 31).
The PlasmaQuant MS is installed and connected by service engineers from Analytik
Jena or by technical personnel authorized by Analytik Jena.
Sample introduction components, including the torch, the nebulizer with spray
chamber and the autosampler have to be installed by the customer during
maintenance operations. Descriptions of these installation procedures are provided in
the sections below.
WARNING
Plasma emits UV radiation and high
-frequency electromagnetic radiation which can
cause serious eye and skin injuries as well as other health problems.
Therefore, the plasma compartment door
must be closed before igniting and operating
the plasma. Safety interlocks must not be bypassed by the user!
Before opening the plasma compartment lid for installation or maintenance work,
always extinguish the plasma from the software.
CAUTION
The torch compartment and interface, including the cones, are still very hot after the
plasma has been extinguished. Contact with these hot surfaces can cause burns.
Therefore, wait 5 minutes after extinguishing the plasma before touching the torch
and th
e cones or wear heat-resistant gloves.