PlasmaQuant MS product family Method optimization
85
If there are not significant improvements in the sensitivities, then check the next
point; otherwise run the ion optics auto optimization again.
Run a plasma auto optimization by selecting a plasma optimization routine during
auto optimization. Then re-run the ion optics optimization.
Double check the instrument setup by running the plasma alignment, mass
calibration, and mass trim (if mass calibration failed), and then re-run the ion
optics auto optimization.