EasyManua.ls Logo

Endress+Hauser Cerabar M - Page 157

Endress+Hauser Cerabar M
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cerabar M, Deltabar M, Deltapilot M Commissioning with the FF configuration program
Endress+Hauser 157
9.8.2 Totalizer
Example:
In this example, the volume flow should be totalized and displayed in the unit m
3
E
3
.
Negative flows should be added to the flow rate.
For a description of the parameters mentioned, see
ä 195, DP_FLOW Transducer Block
ä 208, Analog Input Block.
Totalizer 1 can be reset. Totalizer 2 cannot be reset.
You can use the Display mode/DISPLAY_MAIN_LINE_1_CONTENT parameter ( ä 202)
to specify which measured value should be displayed on the local operation.
Description
1 Calibrate the device in accordance with Section 9.8.1.
2 Open the DP_FLOW Transducer Block and set the block mode to OOS.
3 By means of the Eng. unit total. 1/TOTALIZER_1_UNIT parameter, select a flow unit, here m
3
E
3
for
example.
4 By means of the Totalizer 1 mode/TOTALIZER_1_MODE parameter, specify the totalizing mode for
negative flows, here the "Only negative flow" option for example.
5 Use the Reset Totalizer 1/TOTALIZER_1_RESET parameter to reset to zero.
6 Result:
The Totalizer 1/TOTALIZER_1_STRING_VALUE record parameter, 1/E_TOTALIZER_1_FLOAT totalizer
element displays the totalized volume flow.
7 Set the DP_FLOW Block to "Auto".

Table of Contents

Related product manuals