Cerabar S PMP71 with 4 to 20 mA HART Troubleshooting
Endress+Hauser 49
8.4 Repair
The Endress+Hauser repair concept provides for measuring devices to have a modular design and
that the customer may also carry out repairs ( ä 50, "Spare Parts").
!
Note!
For certified devices, please refer to the "Repair of Ex-certified devices" Section.
For more information on service and spare parts, contact
Endress+Hauser Service. See www.endress.com/worldwide.
8.5 Repair of Ex-certified devices
#
Warning!
When repairing Ex-certified devices, please note the following:
Only specialist personnel or Endress+Hauser may undertake repairs
of certified devices.
Relevant standards, national hazardous area regulations and safety instructions and certificates
must be observed.
Only genuine Endress+Hauser spare parts may be used.
When ordering spare parts, please check the device designation on the nameplate.
Identical parts may only be used as replacements.
Electronic inserts or sensors already in use in a standard device may not be used as spare parts for
a certified device.
Carry out repairs according to the instructions. After repairs, the device must fulfill the
requirements of the specified individual tests.
A certified device may only be converted into another certified variant by
Endress+Hauser.
All repairs and modifications must be documented.