Installation Cerabar S PMP71 with 4 to 20 mA HART
10 Endress+Hauser
3 Installation
3.1 Incoming acceptance, transport and storage
3.1.1 Incoming acceptance
Check the packaging and the contents for damage.
Check the shipment, make sure nothing is missing and that the scope of supply matches your
order.
3.1.2 Transporting to the measuring point
"
Caution!
Please comply with the safety instructions and transport conditions for devices weighing over 18 kg
(39.69 lbs).
Transport the device to the measuring point in the original packaging or at the process connection.
3.1.3 Storage
The device must be stored in a dry, clean place and protected against damage from impact
(EN 837-2).
Storage temperature range:
See Technical Information TI00383P.
3.2 Installation conditions
3.2.1 Dimensions
For dimensions, please refer to the "Mechanical construction" section in TI00383P.
3.3 Installation instructions
!
Note!
Due to the orientation of the Cerabar S, there may be a shift in the zero point, i.e. when the
container is empty or partially full, the measured value does not display zero. You can correct this
zero point shift either directly at the device using the "E" key or by remote operation. See
ä 23, "Function of the operating elements – onsite display not connected" or
ä 35, "Position adjustment".
To ensure optimal readability of the onsite display, it is possible to rotate the housing up to 380°.
ä 14, "Rotating the housing".
Endress+Hauser offers a mounting bracket for installing on pipes or walls.
ä 13, "Wall and pipe-mounting (optional)".