Identification Cerabar S PMP71 with 4 to 20 mA HART
6 Endress+Hauser
2 Identification
2.1 Device designation
2.1.1 Nameplates
!
Note!
• The MWP (maximum working pressure) is specified on the nameplate. This value refers to a
reference temperature of 20°C (68°F) or 100°F (38°C) for ANSI flanges.
• The pressure values permitted at higher temperatures can be found in the following standards:
– EN 1092-1: 2001 Tab. 18
1)
– ASME B 16.5a – 1998 Tab. 2-2.2 F316
– ASME B 16.5a – 1998 Tab. 2.3.8 N10276
– JIS B 2220
• The test pressure corresponds to the overpressure limit (OPL) of the device =
MWP x 1.5
2)
.
• The Pressure Equipment Directive (EC Directive 97/23/EC) uses the abbreviation "PS".
The abbreviation "PS" corresponds to the MWP (maximum working pressure) of the measuring
device.
Aluminum and stainless steel housing (T14)
P01-XMX7Xxxx-18-xx-xx-xx-000
Fig. 1: Nameplate
1 Device name
2 Order code
See the specifications on the order confirmation for the meanings of the individual letters
and digits.
3 Serial number
4 Degree of protection
5 MWP (Maximum working pressure)
6 Symbol: Note: pay particular attention to the data in "Technical Information"!
7 Minimum/maximum span
8 Nominal measuring range
9 Electronic version (output signal)
10 Wetted materials
11 Supply voltage
12 GL symbol for GL marine certificate (optional)
13 SIL symbol for devices with SIL3/IEC 61508 Declaration of Conformity (optional)
14 Approval marks and ID numbers
15 Manufacturer's address
1) With regard to their stability-temperature property, the materials 1.4404 and 1.4435 are grouped together under 13EO
in EN 1092-1 Tab. 18. The chemical composition of the two materials can be identical.
2) The equation does not apply to PMP71 with a 50 bar (750 psi) or a 100 bar (1,500 psi) measuring cell.
Mat.
MWP
Span
Ser.-No.:
Order Code:
U=
1
2
3
4
5
6
7
8
9
10
11
12
13
14
26
23,5
2,5
55
R
1
2,1
-0,1
15