Ecograph T, RSG35 Commissioning
Endress+Hauser 33
When locked, basic settings can be checked but not changed. You can also use a PC to
commission or configure your device.
Device configuration options:
Setup directly at the device
Setup via SD card or USB stick by transferring the parameters stored on it
Setup via Web server using Ethernet
Setup via FieldCare/DeviceCare configuration software using USB interface or Ethernet
9.4.1 Step-by-step: to the first measured value
Procedure and necessary settings:
1. Check the date/time in the main menu under "Setup" and set it if necessary
2. Make settings for the interfaces and communication in the main menu under "Setup
-> Advanced setup -> Communication"
3. Create universal or digital inputs in the main menu under "Setup -> Advanced setup
-> Inputs -> Universal inputs/Digital inputs": Add input: select "Universal input
x" or "Digital input x" with which the input signal should be detected. Then select the
newly created input and configure it
4. Activate relays or analog outputs (optional) in the main menu under "Setup ->
Advanced setup -> Outputs"
5. Assign activated inputs to a group in the main menu under "Setup -> Advanced
setup -> Application -> Signal groups -> Group x"
6. Use "Back" or "ESC" to quit the main menu. The changes made are accepted and saved
The device is in the measured value display mode and displays the measured values.
9.4.2 Step-by-step: set or delete the limit values
Procedure for setting limit values:
1. Open the limit values in the main menu under "Setup -> Advanced setup ->
Application -> Limits"
2. Add a limit value: select "Yes"
3. Select and configure "Limit value x"
4. Use "Back" or "ESC" to quit the main menu. The changes made are accepted and saved
The device is in the measured value display mode and displays the measured values.
Procedure for deleting limit values:
1. Open the limit values in the main menu under "Setup -> Advanced setup ->
Application -> Limits"
2. Delete a limit value: select "Yes"
3. Select the limit value to be deleted from the list
4. Use "Back" or "ESC" to quit the main menu. The changes made are accepted and saved
The device is in the measured value display mode and displays the measured values.
9.4.3 Setup directly at the device
You can access the main menu by pressing the Navigator during operation. Turn the
Navigator to navigate through the available menus. When the desired menu is displayed,
press the Navigator to open the menu.