EasyManua.ls Logo

Endress+Hauser FMR 231 - Mounting position

Endress+Hauser FMR 231
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Mounting position
The ideal mounting position is as follows:
not in the middle of the tank
not above the filling stream
no fittings in beam
avoid vibration, direct high-pressure
cleaning and lateral loads.
min. 30 cm
BA171Y72
A decentral
position avoids
double echoes
BA171Y71
A central
position
strengthens
double echoes
BA171Y73
A position away
from the filling
stream will avoid
interference
echoes
Echos from
fittings are
suppressed
during
calibraatrion
BA171Y74
A position above
the filling stream
will be subject to
interference
echoes
It is not possible
to measure in
tanks with
stirrers
BA171Y76
If the nozzle is too
long, noise will
reduce the signal
quality
BA171Y75
Use either a
longer antenna
or a shorter
nozzle
see page 11 for
details
Chapter 2 Installation Micropilot FMR 231
10 Endress+Hauser