iTEMP TMT82 Identification
Endress+Hauser 9
1
2
3
4
5
6
0044
iTEMP
®
TMT82
Made in Germany 2012 D-87484 Nesselwang
TAGXXXXXXXXXXX
TAGXXXXXXXXXXX
FW: XX.XX.XX Dev. Rev.: X
Order code: TMT82-xxx/xx
Ext. ord. cd.: XXXXXXXXXXX
Ser. no.: XXXXXXXXX
Input: 12-42 V
Current cunsum.: 23,0 mA
1
II3G Ex nA II T6/T5/T4 Ta= -40 ...+ 55/70/85°C
II3G Ex nL IIC T6/T5/T4 Ta= -40 ... +55/70/85°C
EH 08.001X
Install per XA0xxxxT/09/a3/xx.xx
7
A0017924
2 Nameplate of DIN rail transmitter (example, Ex version)
1 Product description and manufacturer ID
2 Order code, extended order code and serial number
3 Power supply and current consumption
4 Approval in hazardous area with number of the relevant Ex documentation (XA...)
5 Firmware version and device version
6 Approval logos
7 2 lines for the TAG name
3.2 Scope of delivery
The scope of delivery of the device comprises:
• Temperature transmitter
• Mounting material (head transmitter)
• Hard copy of multi-language Brief Operating Instructions
• Functional Safety Manual (SIL mode)
• Additional documentation for devices which are suitable for use in the hazardous area
(0 1), such as Safety Instructions (XA), Control or Installation Drawings (ZD).
3.3 Certificates and approvals
The device left the factory in a safe operating condition. The device complies with the
requirements of the standards EN 61 010-1 "Safety Requirements for Electrical Equipment
for Measurement, Control, and Laboratory Use" and with the EMC requirements as per the
IEC/EN 61326 series.
3.3.1 CE/EAC mark, declaration of conformity
The device meets the legal requirements of the EU/EEU guidelines. The manufacturer
confirms that the device is compliant with the relevant guidelines by applying the CE/EAC
mark.
3.3.2
HART
®
protocol certification
The temperature transmitter is registered by the HART
®
FieldComm Group. The device
meets the requirements of the HART Communication Protocol Specifications, Revision 7
(HCF 7.4).