EasyManua.ls Logo

Endress+Hauser Levelflex FMP54 - Page 112

Endress+Hauser Levelflex FMP54
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning (block-based operation) Levelflex FMP51, FMP52, FMP54 FOUNDATION Fieldbus
112 Endress+Hauser
D
L
L
L
F
L
I
UP
D
I
LN
E
R
0%
100%
DK (DC )
2 2
DK DC
1 1
( )
A0011177
 27 Configuration parameters for interface measurements
R = Reference pioint of the measurement D
I
= Distance of interface (Distance from reference point to lower
medium)
E = Empty calibration (= zero point) L
I
= Interface level
F = Full calibration (= span) D
L
= Distance from reference point R to total level
LN = Length of probe L
L
= total level
UP = Thickness of upper medium
Schritt Block Parameter Aktion
1 SETUP (TRDSUP) Distance unit (distance_unit) Select distance unit.
Selection:
1010: m
1013: mm
1018: in
1019: ft
2 SETUP (TRDSUP) Operating mode (operating_mode)
1)
Select 32938: Interface.
3 SETUP (TRDSUP) Tank type (tank_type) Select tank type.
Selection:
32816: Bypass / pipe
33288: Metallic
33302: Coaxial
33432: Twin rope
33433: Twin rod
33437: Rope centering disc metallic
33438: Rod centering disc metallic
33441: Non metallic
33444: Mounted outside
4 SETUP (TRDSUP) Tube diameter (tube_diameter)
2)
Enter the diameter of the bypass or
stilling well.
5 SETUP (TRDSUP) Tank level (tank_level) Select tank level.
Selection:
32919: Fully flooded (typical for
bypass measurements)
33021: Partially filled (typical for
measurements directly in tank)

Table of Contents

Other manuals for Endress+Hauser Levelflex FMP54

Related product manuals