Operating menu Levelflex FMP51, FMP52, FMP54 FOUNDATION Fieldbus
184 Endress+Hauser
1
2
UB
UB
A0013220
1 Suppression of interference echoes at the top end of the probe.
2 Suppression of the level signal in case of a flooded bypass.
UB Upper blocking distance
Level correction
Navigation  Setup → Advanced setup → Interface → Level correction
Description Specify level correction (if required).
User entry –200 000.0 to 200 000.0 %
Factory setting 0.0 %
Additional information The value specified in this parameter is added to the measured total and interface levels
(before linearization).
Manual thickness upper layer
Navigation Setup → Advanced setup → Interface → Man.thick.up.lay
Description Specify the manually determined interface thickness UP (i.e. the thickness of the upper
medium).
User entry 0 to 200 m
Factory setting 0 m