EasyManua.ls Logo

Endress+Hauser Levelflex FMP54 - Page 42

Endress+Hauser Levelflex FMP54
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Mounting Levelflex FMP51, FMP52, FMP54 FOUNDATION Fieldbus
42 Endress+Hauser
Non-metallic vessels
1
2
A0012527
1 Non-metallic vessel
2 Metal sheet or metal flange
To ensure reliable measurements in non-metallic vessels:
Select an instrument version with metal flange (minimum size DN50/2").
Or: mount a metal sheet with a diameter of at least 200 mm (8 in) to the probe at the
process connection. Its orientation must be perpendicular to the probe.
A metallic surface is not required for coax probes.

Table of Contents

Other manuals for Endress+Hauser Levelflex FMP54

Related product manuals