Levelflex FMP51, FMP52, FMP54 FOUNDATION Fieldbus Mounting
Endress+Hauser 53
Mounting the electronics housing
122 (4.8)
52 (2)
86
(3.4)
70
(2.8)
140 (5.5)
158 (6.2)
175 (6.9)
A B
ø42...60
(1.65...2.36)
A0014793
 11 Mounting the electronics housing using the mounting bracket; dimensions: mm (in)
A Wall mounting
B Pipe mounting
Connecting the cable
Required tools:
Open-end wrench 18AF
A
B
C
C
r = 100 (4)
min
r = 100 (4)
min
6 Nm
(4.42 lbf ft)
6 Nm
(4.42 lbf ft)
6 Nm
(4.42 lbf ft)
6 Nm
(4.42 lbf ft)
A0014794
 12 Connecting the cable. There are the following possibilities:
A Angled plug at the probe
B Angled plug at the electronics housing
C Length of the remote cable as ordered