Operation Liquiline CM44P
114 Endress+Hauser
Menu/Setup/Additional functions/Controller 1 ... 2
Function Options Info
XHyst 0.0 to 99.9 %
Factory setting
0.0 %
Width of the hysteresis range of the neutral
zone, relative component of x
n
x
n
x
hyst
e =
w-x
y
The graphic illustrates the manipulated variable (with a pure Pcontroller) over the control deviation e (set
point minus controlled variable). Very low control deviations are set to zero. Control deviations > x
n
are
processed "in the normal way". Via x
hyst
it is possible to configure a hysteresis to suppress oscillations at the
edges.
Clock 0.333 to 100.000 s
Factory setting
1.000 s
Expert setting
Only change the clock time of the controller if
you know exactly what you are doing.
Clock < Twu < 0.5(Tn + Tv)
Max y change rate 0.00 to 1.00
Factory setting
0.40
Limits the change of the output variable
A value of 0.5 permits a maximum manipulated
variable change of 50 % within one second.
Exception behavior Hold=measured value is no longer reliable
Manipulated Variable Selection
Freeze
Fixed value
Factory setting
Freeze
How should the controller react to a measured
value that is no longer reliable?
Freeze
The manipulated variable is frozen at the current
value
Fixed value
Manipulated variable is set to 0 (no dosing)
State Selection
Freeze
Reset
Factory setting
Freeze
Internal controller status
Freeze
No change
Reset
After a hold, the control system starts from
scratch, and settling time takes place again.
Hold as exception Selection
All
None
Factory setting
All
Select: Should the hold trigger the previously
selected exception behaviour, or should it be
ignored?
Outputs Goes to the menu Outputs →  86
Controller assignment view Shows an overview of the inputs and outputs
used
10.6.4 Cleaning programs
L
CAUTION
Cleaning not switched off during calibration or maintenance activities
Risk of injury due to medium or cleaning agent!
If a cleaning system is connected, switch it off before removing a sensor from the
medium.
If you wish to check the cleaning function and have therefore not switched off the
cleaning system, wear protective clothing, goggles and gloves or take other appropriate
measures.