Liquiline CM44P Diagnostics and troubleshooting
Endress+Hauser 143
No. Message Factory settings Tests or remedial action
S
1)
D
2)
F
3)
976 PFM/PWM value high S On Off Pulse frequency modulation: output signal
exceeded/undershot. Measured value outside the
specified range.
Sensor in air
Air pockets in assembly
Incorrect flow to sensor
Sensor fouled
1. Clean sensor
2. Check plausibility.
3. Adjust the PFM configuration.
977 PFM/PWM value low S On Off
978 ChemoClean Failsafe S On On No feedback signal detected within the configured
period.
1. Check application.
2. Check wiring.
3. Extend the duration.
990 Deviation limit F On On Redundancy: limit value of percentage deviation
exceeded
991 CO2 conc. range F On On CO
2
concentration (degassed conductivity) outside
the measuring range
992 pH calculation range F On On pH calculation outside the measuring range
993 rH calculation range F On On rH calculation outside the measuring range
994 Difference conduct. F On On Dual conductivity outside the measuring range
1) Status signal
2) Diagnostic message
3) Failure current
12.6.2 Sensor-specific diagnostic messages
The following abbreviations for the various sensor types are used in the table:
P ... pH/ORP (general, applies for all pH sensors)
P (glass) ... only applies for glass electrodes
P (ISFET) ... only applies for ISFET sensors
C ... conductivity (general, applies for all conductivity sensors)
C (cond.) ... only applies for sensors with conductive measurement of conductivity
C (ind.) ... only applies for sensors with inductive measurement of conductivity
O ... oxygen (general, applies for all oxygen sensors)
O (opt.) ... only applies for optical oxygen sensors
O (amp.) ... only applies for amperometric oxygen sensors
N ... nitrate sensors
T ... turbidity and solids sensors
S ... SAC sensors
U ... interface sensors
I ... ion-selective sensors
DI ... disinfection sensors
Phot ... analog photometer sensors
No. Message Factory settings Sensor
type
Tests or remedial action
S
1)
D
2)
F
3)
002 Sensor unknown F On On All
Replace sensor.
004 Sensor defective F On On All