Liquiline CM44P Electrical connection
Endress+Hauser 39
Connecting photometer sensors to PEM module
Sensor Cable color PEM terminal Assignment
OUSAF11
OUSAF12
YE (thick) P+ Lamp voltage +
YE (thin) S+ Recording lamp voltage +
BK (thin) S- Recording lamp voltage -
BK (thick) P- Lamp voltage -
RD A (1) Sensor +
BK
1)
/ WH
2)
C(1) Sensor -
GY SH (1) Shield
OUSAF21
OUSAF22
OUSTF10
OUSAF44
YE (thick) P+ Lamp voltage +
YE (thin) S+ Recording lamp voltage +
BK (thin) S- Recording lamp voltage -
BK (thick) P- Lamp voltage -
RD A (1) Measuring detector sensor +
BK C(1) Measuring detector sensor -
GY SH (1) Measuring detector screening
WH A (2) Sensor reference +
GN C(2) Sensor reference -
GY SH (2) Reference screening
OUSAF46
2 PEM modules
necessary
PEM module 1
YE (thick) P+ Lamp voltage +
YE (thin) S+ Recording lamp voltage +
BK (thin) S- Recording lamp voltage -
BK (thick) P- Lamp voltage -
RD A (1) Measuring detector sensor +
BK C(1) Measuring detector sensor -
GY SH (1) Measuring detector screening
WH (lamp) A (2) Sensor reference +
GN (lamp) C(2) Sensor reference -
GY (lamp) SH (2) Reference screening
PEM module 2
WH A (1) Measuring detector sensor +
GN C(1) Measuring detector sensor -
GY SH (1) Measuring detector screening
RD (lamp) A (2) Sensor reference +
BK (lamp) C(2) Sensor reference -
GY (lamp) SH (2) Reference screening