Liquiline CM44P Operation
Endress+Hauser 87
Menu/Setup/Outputs/Current output x:y
1)
Function Options Info
Hold behavior Selection
• Freeze last value
• Fixed value
• Ignore
Factory setting
Depends on the
channel:output
Freeze last value
The device freezes the last current value.
Fixed value
You define a fixed current value that is output at
the output.
Ignore
A hold does not affect this current output.
Hold current
Hold behavior = Fixed value
0.0 to 23.0 mA
Factory setting
22.0 mA
‣
Specify which current should be output at
this current output in the hold state.
1) x:y = slot:output number
Measured value depending on the Source of data
Source of data Measured value
Photometer Selection
• Measured value
• 2nd measurement value
• Raw meas. current
• Raw ref. current
• Lamp current
• Lamp voltage
• Raw meas. value
• Raw 2nd meas. value
pH Glass Selection
• Raw value mV
• pH
• Temperature
pH ISFET
ORP Selection
• Temperature
• ORP mV
• ORP %
Oxygen (amp.) Selection
• Temperature
• Partial pressure
• Concentration liquid
• Saturation
• Raw value nA
(only Oxygen (amp.))
• Raw value μs
(only Oxygen (opt.))
Oxygen (opt.)
Cond i Selection
• Temperature
• Conductivity
• Resistance
(only Cond c)
• Concentration
(only Cond i and Cond c 4-pol)
Cond c
Cond c 4-pol
Disinfection Selection
• Temperature
• Sensor current
• Concentration
ISE Selection
• Temperature
• pH
• Ammonium
• Nitrate
• Potassium
• Chloride