Operation Liquiline CM44P
92 Endress+Hauser
Outputting the status of a cleaning function
Function = Cleaning
Function Options Info
Assignments Selection
None
Depends on the type of
cleaning
Factory setting
None
Here you can specify how a cleaning function
should be displayed for the relay.
You have the following options depending on the
cleaning program that is selected (Menu/Setup/
Additional functions/Cleaning:
Cleaning type = Standard clean
Cleaning 1 - Water, Cleaning 2 - Water,
Cleaning 3 - Water, Cleaning 4 - Water
Cleaning type = Chemoclean
Cleaning 1 - Water, Cleaning 1 - Cleaner,
Cleaning 2 - Water, Cleaning 2 - Cleaner,
Cleaning 3 - Water, Cleaning 3 - Cleaner,
Cleaning 4 - Water, Cleaning 4 - Cleaner
Cleaning type = Chemoclean Plus
4x Cleaning 1 - %0V, 4x Cleaning 2 - %0V
1)
Hold behavior Selection
Freeze last value
Fixed value
Ignore
Factory setting
Ignore
Freeze last value
The device freezes the last measured value.
Fixed value
You define a fixed measured value that is output
at the output.
Ignore
A hold has no effect.
1) %0V is variable text which you can assign in Menu/Setup/Additional functions/Cleaning/Chemoclean
Plus/Output label 1 ... 4.
Formula
Function = Controller
Function Options Info
Operating mode Selection
PWM
PFM
Factory setting
PWM
PWM=pulse width modulation
PFM=pulse frequency modulation→  90
Source of data Selection
List of available
formulas
Maximum of 8 formulas
Mathematical functions of type Formula must be
available.
Select the formula which is to serve as a data
source.
Range lower value 0 to 9999
Range upper value Range lower value to
9999
Cycle duration
Operating mode = PWM
Shortest turn-on time to
999.0 s
Factory setting
10.0 s
Specify the cycle duration within which the
duty cycle should be varied (only PWM).
The settings for Cycle duration and Shortest turn-on time affect one another. The following applies:
Cycle durationShortest turn-on time.
Shortest turn-on time
Operating mode = PWM
0.3 s to Cycle duration
Factory setting
0.3 s
Pulses that are shorter than this limit value are
not output in order to conserve the actuator.