Micropilot FMR60B PROFINET with Ethernet-APL
12 Endress+Hauser
PTFE Drip-off antenna, 50 mm (2 in) in vessel with agitator
Media group Measuring range
A0 (ε
r
1.2 to 1.4) 2 m (7 ft)
A (ε
r
1.4 to 1.9) 4 m (13 ft)
B (ε
r
1.9 to 4) 7 m (23 ft)
C (ε
r
4 to 10) 15 m (49 ft)
D (ε
r
>10) 25 m (82 ft)
Operating frequency
Approx. 80 GHz
Up to 8 devices can be installed in a tank without the devices mutually influencing one another.
Transmission power
• Peak power: 6.3 mW
• Average output power: 63 µW
Output
PROFINET-APL
PROFINET with Ethernet-APL
10BASE-T1L, 2-wire 10 Mbit/s
Signal on alarm
Local display
Status signal (in accordance with NAMUR Recommendation NE 107):
Plain text display
Operating tool via service interface (CDI)
Status signal (in accordance with NAMUR Recommendation NE 107):
Plain text display
Operating tool via PROFINET with Ethernet-APL
• According to "Application layer protocol for decentralized periphery", Version 2.4
• Diagnostics according to PROFINET PA Profile 4.02
Linearization
The linearization function of the device allows the conversion of the measured value into any unit of
length, weight, flow or volume.
Pre-programmed linearization curves
Linearization tables for calculating the volume in the following vessels are preprogrammed into the
device:
• Pyramid bottom
• Conical bottom
• Angled bottom
• Horizontal cylinder
• Sphere
Other linearization tables of up to 32 value pairs can be entered manually.
PROFINET with Ethernet-
APL
Protocol Application layer protocol for decentral device periphery and distributed
automation, Version 2.4
Communication type Ethernet Advanced Physical Layer 10BASE-T1L
Conformance Class Conformance Class B
Netload Class Netload Class II