Endress+Hauser Micropilot FMR60B - Storage Temperature; Climate Class; Installation Height as Per IEC61010-1

Endress+Hauser Micropilot FMR60B
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR60B PROFINET with Ethernet-APL
36 Endress+Hauser
316L housing, hygiene; process temperature –40 to +130 °C (–40 to +266 °F)
T
a
T
p
P2
P1
P3
P4
P5
A0032024
 38 316L housing, hygiene; process temperature –40 to +130 °C (–40 to +266 °F)
P1 = T
p
: –40 °C (–40 °F) |  T
a
: +76 °C (+169 °F)
P2 = T
p
: +76 °C (+169 °F) |  T
a
: +76 °C (+169 °F)
P3 = T
p
: +130 °C (+266 °F) |  T
a
: +55 °C (+131 °F)
P4 = T
p
: +130 °C (+266 °F) |  T
a
: –40 °C (–40 °F)
P5 = T
p
: –40 °C (–40 °F) |  T
a
: –40 °C (–40 °F)
316L housing, hygiene; process temperature –40 to +150 °C (–40 to +302 °F)
T
a
T
p
P2
P1
P3
P4
P5
A0032024
 39 316L housing, hygiene; process temperature range: –40 to +150 °C (–40 to +302 °F)
P1 = T
p
: –40 °C (–40 °F) |  T
a
: +76 °C (+169 °F)
P2 = T
p
: +76 °C (+169 °F) |  T
a
: +76 °C (+169 °F)
P3 = T
p
: +150 °C (+302 °F) |  T
a
: +41 °C (+106 °F)
P4 = T
p
: +150 °C (+302 °F) |  T
a
: –40 °C (–40 °F)
P5 = T
p
: –40 °C (–40 °F) |  T
a
: –40 °C (–40 °F)
316L housing, hygiene; process temperature –40 to +200 °C (–40 to +392 °F)
T
a
T
p
P2
P1
P3
P4
P5
A0032024
 40 316L housing, hygiene; process temperature –40 to +200 °C (–40 to +392 °F)
P1 = T
p
: –40 °C (–40 °F) |  T
a
: +76 °C (+169 °F)
P2 = T
p
: +76 °C (+169 °F) |  T
a
: +76 °C (+169 °F)
P3 = T
p
: +200 °C (+392 °F) |  T
a
: +32 °C (+90 °F)
P4 = T
p
: +200 °C (+392 °F) |  T
a
: –40 °C (–40 °F)
P5 = T
p
: –40 °C (–40 °F) |  T
a
: –40 °C (–40 °F)
Storage temperature
Without LCD display: –40 to +90 °C (–40 to +194 °F)
With LCD display: –40 to +85 °C (–40 to +185 °F)
Climate class
DIN EN 60068-2-38 (test Z/AD)
Installation height as per
IEC61010-1 Ed.3
Generally up to 5 000 m (16 404 ft) above sea level