Micropilot FMR60B PROFINET with Ethernet-APL
18 Endress+Hauser
Depending on the type of port (DC power supply, input/output port) different testing levels according
to IEC / DIN EN 61326-1 against transient overvoltages (Surge) are applied (IEC / DIN EN
61000-4-5 Surge):
Test level on DC power ports and input/output ports is 1 000 V line to earth
Devices with optional overvoltage protection
Spark-over voltage: min. 400 V
DC
Tested according to IEC / DIN EN 60079-14 sub chapter 12.3 (IEC / DIN EN 60060-1 chapter 7)
Nominal discharge current: 10 kA
NOTICE
Device could be destroyed
Always ground device with integrated overvoltage protection.
Overvoltage category
Overvoltage category II
Performance characteristics
Reference operating
conditions
Temperature = +24 °C (+75 °F) ±5 °C (±9 °F)
Pressure = 960 mbar abs. (14 psia) ±100 mbar (±1.45 psi)
Humidity = 60 % ±15 %
Reflector: metal plate with diameter ≥ 1 m (40 in)
No major interference reflections inside the signal beam
Maximum measured error Reference accuracy
Accuracy
The accuracy is the sum of the non-linearity, non-repeatability and hysteresis.
Measuring distance up to 0.8 m (2.62 ft): max. ±4 mm (±0.16 in)
Measuring distance > 0.8 m (2.62 ft): ±1 mm (±0.04 in)
Non-repeatability
Non-repeatability is already included in the accuracy.
≤ 1 mm (0.04 in)
If conditions deviate from the reference operating conditions, the offset/zero point that results
from the installation conditions can be up to ±4 mm (±0.16 in). This additional offset/zero
point can be eliminated by entering a correction (Level correction parameter) during
commissioning.