EasyManua.ls Logo

Endress+Hauser Micropilot FMR60B - Access to the operating menu via the operating tool

Endress+Hauser Micropilot FMR60B
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR60B PROFINET with Ethernet-APL
Endress+Hauser 37
Degree of protection
Test as per IEC 60529 and NEMA 250-2014
Housing
IP66/68, NEMA TYPE 4X/6P
IP68 test condition: 1.83 m under water for 24 hours.
Cable entries
Gland M20, plastic, IP66/68 NEMA TYPE 4X/6P
Gland M20, nickel-plated brass, IP66/68 NEMA TYPE 4X/6P
Gland M20, 316L, IP66/68 NEMA TYPE 4X/6P
Gland M20, hygiene, IP66/68/69 NEMA Type 4X/6P
Thread M20, IP66/68 NEMA TYPE 4X/6P
Thread G1/2, IP66/68 NEMA TYPE 4X/6P
If the G1/2 thread is selected, the device is delivered with an M20 thread as standard and a G1/2
adapter is included with the delivery, along with the corresponding documentation
Thread NPT1/2, IP66/68 NEMA TYPE 4X/6P
M12 plug
When housing is closed and connecting cable is plugged in: IP66/67 NEMA TYPE 4X
When housing is open or connecting cable is not plugged in: IP20, NEMA TYPE 1
NOTICE
M12 plug: Loss of IP protection class due to incorrect installation!
The degree of protection only applies if the connecting cable used is plugged in and screwed
tight.
The degree of protection only applies if the connecting cable used is specified according to
IP66/67 NEMA 4X.
The protection classes are only maintained if the dummy cap is used or the cable is connected.
Vibration resistance
DIN EN 60068-2-64 / IEC 60068-2-64 for 5 to 2 000 Hz: 1.5 (m/s
2
)
2
/Hz
Electromagnetic
compatibility (EMC)
Electromagnetic compatibility as per EN 61326 series and NAMUR recommendation EMC (NE21)
Maximum measured error during EMC testing: < 0.5 % of the current digital measured value
For more details refer to the EU Declaration of Conformity.
Process
Process pressure range
L
WARNING
The maximum pressure for the device depends on the lowest-rated component with regard to
pressure (components are: process connection, optional mounted parts or accessories).
Only operate the device within the specified limits for the components!
MWP (Maximum Working Pressure): The MWP is specified on the nameplate. This value refers
to a reference temperature of +20 °C (+68 °F) and may be applied to the device for an unlimited
time. Note temperature dependence of MWP. For flanges, refer to the following standards for
the permitted pressure values at higher temperatures: EN 1092-1 (with regard to their stability/
temperature property, the materials 1.4435 and 1.4404 are grouped together under EN 1092-1;
the chemical composition of the two materials can be identical), ASME B16.5, JIS B2220 (the
latest version of the standard applies in each case). MWP data that deviate from this are
provided in the relevant sections of the Technical Information.
The Pressure Equipment Directive (2014/68/EU) uses the abbreviation PS . This corresponds to
the maximum working pressure (MWP) of the device.
The following tables show the dependencies between the seal material, process temperature (T
P
)
and process pressure range for each process connection that can be selected for the antenna used.

Table of Contents

Other manuals for Endress+Hauser Micropilot FMR60B

Related product manuals