Micropilot FMR60B PROFINET with Ethernet-APL
60 Endress+Hauser
Ordering information
Detailed ordering information is available from your nearest sales organization
www.addresses.endress.com or in the Product Configurator at www.endress.com:
1. Select the product using the filters and search field.
2. Open the product page.
3. Select Configuration.
Product Configurator - the tool for individual product configuration
Up-to-the-minute configuration data
Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
Automatic verification of exclusion criteria
Automatic creation of the order code and its breakdown in PDF or Excel output format
Ability to order directly in the Endress+Hauser Online Shop
Calibration Factory calibration certificate
The calibration points are spread evenly over the measuring range (0 to 100 %). The Empty
calibration E and Full calibration F must be specified to define the measuring range. If this
information is missing, antenna-dependent default values are used instead.
R
A
F
E
100%
0%
A0032643
R Reference point of measurement
A Minimum distance between reference point R and 100% mark
E Empty calibration
F Full calibration
Measuring range restrictions
The following restrictions must be considered when selecting E and F:
Minimum distance between reference point R and 100% mark
A ≥ 400 mm (16 in)
Minimum span
F ≥ 45 mm (1.77 in)
Maximum value for Empty calibration
E ≥ 450 mm (17.72 in) (maximum 30 m (98 ft))
Calibration takes place under reference conditions .
The selected values for Empty calibration and Full calibration are only used to create the
factory calibration certificate. Afterwards, the values are reset to the default values specific
for the antenna. If values other than the default values are required, they must be ordered as
a customized empty/full calibration.
Product Configurator → Optional → Service → Customized empty/full calibration
Service
Services that can be selected via the product structure in the Product Configurator include:
Cleaned of oil+grease (wetted)
PWIS-free (paint-wetting impairment substances)
ANSI Safety Red coating, coated housing cover
Set damping
Set max. alarm current