26 Endress+Hauser
The material thickness can be a full multiple of the optimum thickness (table); it is important to
note, however, that the microwave transparency decreases significantly with increasing material
thickness.
Optimum material thickness
Material Optimum material thickness
PE; ε
r
 2.3 1.25 mm (0.049 in)
PTFE; ε
r
 2.1 1.30 mm (0.051 in)
PP; ε
r
 2.3 1.25 mm (0.049 in)
Perspex; ε
r
 3.1 1.10 mm (0.043 in)
Container with heat insulation
1
2
A0046566
If process temperatures are high, the device should be included in the usual container insulation
system (2) to prevent the electronics from heating as a result of thermal radiation or convection. The
rib structure (1) must not be insulated.
Environment
Ambient temperature range
The following values apply up to a process temperature of +85 °C (+185 °F). At higher process
temperatures, the permitted ambient temperature is reduced.
Without LCD display:
Standard: –40 to +85 °C (–40 to +185 °F)
With LCD display: –40 to +85 °C (–40 to +185 °F) with limitations in optical properties such as
display speed and contrast for example. Can be used without limitations up to
–20 to +60 °C (–4 to +140 °F)
If operating outdoors in strong sunlight:
Mount the device in the shade.
Avoid direct sunlight, particularly in warm climatic regions.
Use a weather protection cover (see accessories).
Ambient temperature limits
The permitted ambient temperature (T
a
) depends on the selected housing material (Product
Configurator → Housing; Material →) and the selected process temperature range (Product
Configurator → Application →).
In the event of temperature (T
p
) at the process connection, the permitted ambient temperature (T
a
)
is reduced.
The following information only takes functional aspects into consideration. Additional
restrictions may apply for certified device versions.