Micropilot FMR60B PROFINET with Ethernet-APL
Endress+Hauser 21
Orientation Internal vessel fittings
α
A0031777
Avoid internal fittings (point level switches, temperature sensors, struts, vacuum rings, heating coils,
baffles etc.) inside the signal beam. Pay attention to the beam angle α.
Avoiding interference echoes
A0031813
Metal deflector plates, installed at an angle to scatter the radar signals, help prevent interference
echoes.
Vertical alignment of antenna axis
Align the antenna so that it is perpendicular to the product surface.
The maximum reach of the antenna can be reduced, or additional interference signals can
occur, if the antenna is not installed perpendicular to the product.