Micropilot FMR60B PROFINET with Ethernet-APL
Endress+Hauser 67
Material of single compartment housing, remote display
• Aluminum
• Plastic
Degree of protection:
• IP68 / NEMA 6P
• IP66 / NEMA 4x
Connecting cable:
• Connecting cable (option) up to 30 m (98 ft)
• Customer-supplied standard cable up to 60 m (197 ft)
Recommendation: EtherLine®-P CAT.5e from LAPP.
Specification of customer-supplied connecting cable
Push-in CAGE CLAMP®, connection technology, push actuation
• Conductor cross-section:
• Solid conductor 0.2 to 0.75 mm
2
(24 to 18 AWG)
• Fine-stranded conductor 0.2 to 0.75 mm
2
(24 to 18 AWG)
• Fine-stranded conductor; with insulated ferrule 0.25 to 0.34 mm
2
• Fine-stranded conductor; without insulated ferrule 0.25 to 0.34 mm
2
• Stripping length 7 to 9 mm (0.28 to 0.35 in)
• Outer diameter: 6 to 10 mm (0.24 to 0.4 in)
• Maximum cable length: 60 m (197 ft)
Ambient temperature:
• –40 to +80 °C (–40 to +176 °F)
• Option: –50 to +80 °C (–58 to +176 °F)
Gas-tight feedthrough
Chemically inert glass feedthrough, which prevents gases from entering the electronics housing.
Can optionally be ordered as "Accessory mounted" via the product structure.
Field Xpert SMT70
Universal, high-performance tablet PC for device configuration in Ex Zone 2 and non-Ex areas
For details, see "Technical Information" TI01342S
DeviceCare SFE100
Configuration tool for HART, PROFIBUS and FOUNDATION Fieldbus field devices
Technical Information TI01134S
FieldCare SFE500
FDT-based plant asset management tool
It can configure all smart field units in your system and helps you manage them. By using the status
information, it is also a simple but effective way of checking their status and condition.
Technical Information TI00028S
Documentation
For an overview of the scope of the associated Technical Documentation, refer to the following:
• Device Viewer (www.endress.com/deviceviewer): Enter the serial number from the
nameplate
• Endress+Hauser Operations app: Enter serial number from nameplate or scan matrix code on
nameplate.