Micropilot FMR60B PROFINET with Ethernet-APL
Endress+Hauser 9
Integrated antenna, PEEK, 40 mm (1.5 in) in storage vessel
Media group Measuring range
A0
r
1.2 to 1.4) 3 m (10 ft)
A
r
1.4 to 1.9) 6 m (20 ft)
B
r
1.9 to 4) 11 m (36 ft)
C
r
4 to 10) 15 m (49 ft)
D
r
>10) 22 m (72 ft)
Encapsulated antenna, PVDF, 40 mm (1.5 in) in storage vessel
Media group Measuring range
A0
r
1.2 to 1.4) 7 m (23 ft)
A
r
1.4 to 1.9) 15 m (49.2 ft)
B
r
1.9 to 4) 30 m (98.4 ft)
C
r
4 to 10) 40 m (131 ft)
D
r
>10) 40 m (131 ft)
PTFE Drip-off antenna, 50 mm (2 in) in storage vessel
Media group Measuring range
A0
r
1.2 to 1.4) 7 m (23 ft)
A
r
1.4 to 1.9) 12 m (39 ft)
B
r
1.9 to 4) 23 m (75 ft)
C
r
4 to 10) 40 m (131 ft)
D
r
>10) 50 m (164 ft)
Measurement in buffer vessel
Buffer vessel - measuring conditions
Moving medium surface (e.g. permanent free filling from above, mixing jets)

Table of Contents

Other manuals for Endress+Hauser Micropilot FMR60B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Micropilot FMR60B and is the answer not in the manual?

Endress+Hauser Micropilot FMR60B Specifications

General IconGeneral
BrandEndress+Hauser
ModelMicropilot FMR60B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals