Operating Instructions OXY5500 Gas Analyzer
58 Endress+Hauser
Register Name Address Size Variable Type Description Write Access
Baud Rate Minimum
RS-232
4117 2 Int32 Minimum code for baud rate No
Baud Rate Maximum
RS-232
4119 2 Int32 Maximum code for baud rate No
Parity RS-232 4121 2 Int32 Parity for the RS-232 output where:
0x00 = Even Parity
0x01 = Odd Parity
0x02 = No Parity2
Yes
4-20mA Port1
Output Interface
4359 2 Int32 Code for 4-20mA Port1 Output Mode where:
0x00 = Off
0x01 = Fixed
0x02 = Linear
0x04 = Bilinear
Yes
4-20mA Port1
Output Channel
4363 2 Int32 Code for 4-20mA Port1 Output Interface where:
0x01 = Oxygen
0x20 = Temperature
Yes
4-20mA Port1 Low
Value
4377 2 Float The 4mA output value. Yes
4-20mA Port1 Mid
Value
4379 2 Float
The 12mA output value, which is used in bilinear
mode only.
Yes
4-20mA Port1 High
Value
4381 2 Float The 20mA output value. Yes
4-20mA Port1 Fixed
Value
4383 2 Float
In fixed output mode, this value is applied to the
output. Unit is mA.
Yes
4-20mA Port1 Error
Trigger Level Value
4389 2 Int32 Output current in the event of an error, where:
0x00 = 22mA
0x01 = 2mA
Yes
4-20mA Port1
Calibration Values
4329 8 Float
2 calibration values for a low point and a high
point (each with a reference value and device
output).
4-20mA Port2
Output Interface
4945 2 Int32 Code for 4-20mA Port1 Output Mode where:
0x00 = off
0x01 = fixed
0x02 = linear
0x04 = bilinear
Yes
4-20mA Port2
Output Channel
4949 2 Int32/ Code for 4-20mA Port1 Output Interface where:
0x01 = Oxygen
Yes
4-20mA Port2 Low
Value
4963 2 Float The 4mA output value. Yes
4-20mA Port2 Mid
Value
4965 2 Float
The 12mA output value is only used in bilinear
mode.
Yes
4-20mA Port2 High
Value
4967 2 Float The 20mA output value. Yes
4-20mA Port2 Fixed
Value
4969 2 Float
In fixed output mode, this value is applied to the
output.
Yes
4-20mA Port2 Error
Trigger Level Value
4975 2 Int32 Output current in the event of an error, where:
0x00 = 22mA
Yes
4-20mA Port2
Calibration Values
4979 8 Float
Two calibration values for a low point and a high
point, each with a reference value and device
output.