OXY5500 Gas Analyzer Operating Instructions
Endress+Hauser 59
Register Name Address Size Variable Type Description Write Access
4-20mA Input
Interface
5633 2 Int32 This register is reserved for future use. Yes
4-20mA Input
Channel
5637 2 Int32 Code for 4-20mA Port1 Output Interface where:
0x02 = Pressure.3
No
4-20mA Input Low
Value
5651 2 Float The 4mA corresponding input value. Yes
4-20mA Input Mid
Value
5653 2 Float
The 12mA input value is only used in bilinear
mode.
Yes
4-20mA Input High
Value
5655 2 Float The 20mA input value. Yes
4-20mA Input Fixed
Value
5657 2 Float This register is reserved for future use. Yes
4-20mA Input Error
Trigger Level Value
5663 2 Float This register is reserved for future use. Yes
4-20mA Input
Calibration Values
5667 8 Float
Two calibration values for a low and a high point,
each with reference value and device output.
Yes
Measurement
Values
4895 14 Mixed Refer to Measurement values →  for details. No
Sensor Constant f1 4911 2 Float Sensor constant f1. Allowed range: 0.000 to 9.999 Yes
Sensor Constant
dPhi1
4913 2 Float
Sensor constant dPhi1. Allowed range: -9.99999
to +9.99999
Yes
Sensor Constant
dPhi2
4917 2 Float
Sensor constant dPhi2. Allowed range: -9.99999
to +9.99999
Yes
Sensor Constant
dKSV1
4919 2 Float
Sensor constant dKSV1. Allowed range: -9.99999
to +9.99999
Yes
Sensor constant
DKSV2
4921 2 Float
Sensor constant dKSV2. Allowed range: -9.99999
to +9.99999
Yes
Sensor Constant m 4923 2 Float
Sensor constant m. Allowed range: 0.00 to
+999.99
Yes
Sensor Type 4925 2 Int32 Sensor Type where:
0x00 = OP-3
0x01 = OP-6
0x02 = OP-94
Yes
Manual
Temperature
Compensation
5611 1 Boolean
Activates the temperature measurement of the
Pt100 sensor by setting and uses the manual
temperature value by deleting this Boolean
register. After writing this register, the manual
temperature value has to be set (register 2411).
Yes
Cal0 5521 2 Float
Calibration value: Phase shift of low oxygen
calibration point (Default: 59.9).
Yes
T0 5523 2 Float
Calibration value: Temperate at the low oxygen
calibration point in °C (Default: 20.0).
Yes
O2-2nd 5527 2 Float
Calibration value: Oxygen concentration of the
high oxygen calibration point in the unit defined in
register 5535 (O2-2nd Unit).
Yes
Cal-2nd 5529 2 Float
Calibration value: Phase shift of high oxygen
calibration point (Default: 26.3).
Yes
T2nd 5531 2 Float
Calibration value: Temperature at the high
calibration point in °C.
Yes
pATM 5533 2 Float
Calibration value: Pressure at the high oxygen
calibration in mbar.
Yes