4 Wiring PROline Promass 80
24 Endress+Hauser
4.1.2 Cable specifications
The specifications of the cable connecting the transmitter and the sensor of the remote
version are as follows:
• 6 x 0.38 mm
2
PVC cable with common shield and individually shielded cores.
• Conductor resistance: ≤ 50 Ω/km
• Capacitance: core/shield: ≤ 420 pF/m
• Cable length: max. 20 m
• Permanent operating temperature: max. +105 °C