PROline Promass 80 9 Trouble-shooting
Endress+Hauser 61
9 Trouble-shooting
9.1 Trouble-shooting instructions
Always start trouble-shooting with the checklists below, if faults occur after start-up or
during operation. The routine takes you directly to the cause of the problem and the
appropriate remedial measures.
Check the display
No display visible and no
output signals present.
1. Check the power supply terminals 1, 2
2. Check the power line fuse
Page 77
85...260 V AC: 0.8 A slow-blow / 250 V
20...55 V AC and 16...62 V DC: 2 A slow-blow / 250 V
3. Measuring electronics defective
order spare parts Page 72
No display visible, but
output signals are present.
1. Check whether the ribbon cable connector of the display module is
correctly plugged into the amplifier board
Page 74, 76
2. Display module defective
order spare parts Page 72
3. Measuring electronics defective
order spare parts Page 72
Display texts are in a foreign
language.
Switch off power supply. Press and hold down both the +/
keys and
switch on the measuring device. The display text will appear in English
(default) and is displayed at maximum contrast.
Measured value indicated,
but no signal at the current
or pulse output
Measuring electronics defective
order spare parts Page 72
Error messages on display
Errors which occur during commissioning or measuring operation are displayed immediately.
Error messages consist of a variety of icons. The meanings of these icons are as follows:
Error type: S = system error, P = process error
Error message type:
$ = fault message, ! = notice message
MEDIUM INHOM. = error designation, e.g. fluid is not homogeneous
03:00:05 = duration of error occurrence (in hours, minutes and seconds)
# 702 = error number
" Caution:
Also observe the information on Page 34 ff.!
The measuring system interprets simulations and positive zero return as system errors, but displays
them as notice message only.
Error number:
No. 001 – 400
No. 601 – 699
System error (device error) has occurred
Page 62
Error number:
No. 500 – 600
No. 700 – 750
Process error (application error) has occurred
Page 67
Other error (without error message)
Some other error has
occurred.
Diagnosis and rectification Page 69