6 Commissioning PROline Promass 80
56 Endress+Hauser
6.2.6 Current outputs 2: active/passive
The current outputs are configured as “active” or “passive” by means of various jumpers
on the current input submodule.
#
Warning:
Risk of electric shock. Exposed components carry dangerous voltages. Make sure that
the power supply is switched off before you remove the cover of the electronics
compartment.
1. Switch off power supply.
2. Remove the I/O board
→ Page 73, 75
3. Set the jumpers in accordance with Fig. 26.
" Caution:
Risk of destroying the measuring device. Set the jumpers exactly as shown in
Fig. 26. Incorrectly set jumpers can cause overcurrents that would destroy either
the measuring device or external devices connected to it.
4. Installation of the I/O board is the reverse of the removal procedure.
Fig. 26: Configuring current outputs with the aid of jumpers (I/O board)
1 Current output 1 with HART
1.1 Active current output (default)
1.2 Passive current output
2 Current output 2 (plug-in module)
2.1 Active current output (default)
2.2 Passive current output
F06-x3xxxxxx-16-xx-06-xx-001