Diagnostics and troubleshooting Proline Promass I 300 PROFINET
214 Endress+Hauser
Diagnostic information Remedy instructions
No. Short text
942 API density out of specification 1. Check process density with selected API commodity group
2. Check API related parameters
Measured variable status [from the factory]
1)
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal S
Diagnostic behavior Warning
Influenced measured variables
Mass flow
1) Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
Diagnostic information Remedy instructions
No. Short text
943 API pressure out of specification 1. Check process pressure with selected API commodity group
2. Check API related parameters
Measured variable status [from the factory]
1)
Quality Bad
Quality substatus Maintenance alarm
Coding (hex) 0x24 to 0x27
Status signal S
Diagnostic behavior Warning
Influenced measured variables
Oil density
Water density
GSV flow
GSV flow alternative
Mass flow
Oil mass flow
Water mass flow
NSV flow
NSV flow alternative
External pressure
S&W volume flow
Reference density alternative
Corrected volume flow
Oil corrected volume flow
Water corrected volume flow
Oil volume flow
Water volume flow
Water cut
1) Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.