4. Attach the sensor housing to the sensor holder by locking the bracket into place.
A0043378
44 Fastening the sensor housing
5. Connect the sensor cable to the adapter cable.
This completes the mounting procedure. The sensors can be connected to the transmitter
via the connecting cables.
• To ensure good acoustic contact, the visible measuring pipe surface must be clean and free
from flaking paint and/or rust.
• If necessary, the holder and sensor housing can be secured with a screw/nut or a lead seal
(not supplied).
• The bracket can only be released using an auxiliary tool e.g. screwdriver).
Installing sensors – medium/large nominal diameters DN 50 to 4000 (2 to 160")
Installation for measurement via 1 traverse
Requirements
• The installation distance and wire length are known → 44
• Strapping bands are pre-assembled
Material
The following material is required for mounting:
• Two strapping bands incl. mounting bolts and centering plates where necessary (already pre-
assembled → 46, → 47)
• Two measuring wires, each with a cable lug and a fixer to fix the strapping bands
• Two sensor holders
• Coupling medium (coupling pad or coupling gel) for the acoustic connection between the sensor
and pipe
• Two sensors incl. connecting cables
Installation is unproblematic up to DN 400 (16"); as of DN 400 (16") check the distance and
angle (180°, ±5°) diagonally with the wire length.
Procedure for using measuring wires:
1. Prepare the two measuring wires: arrange the cable lugs and fixer such that the distance they
are apart corresponds to the wire length (SL). Screw the fixer onto the measuring wire.
1 2 3 4 5 6 7 8 9 10 11
SL
A0043379
45 Fixer and cable lugs at a distance that corresponds to the wire length (SL)