Procedure:
1. Set the distance between the sensors to the value determined for the sensor distance. Press the
movable sensor down slightly to move it.
b
a
A0043376
 42 Distance between sensors as per the installation distance →  44
a Sensor distance (back of sensor must touch the surface)
b Sensor contact surfaces
2. Stick the coupling pad under the sensor onto the measuring pipe. Alternatively, coat the
contact surfaces of the sensor (b) evenly with coupling gel (approx.
0.5 to 1 mm (0.02 to 0.04 in)).
3. Position the sensor housing on the sensor holder.
A0043377
 43 Positioning the sensor housing